Commissioning
27
28.3
• Press OK for approx. 1 second.
The current bath temperature will be shown on
LED temperature display and the temperature
control is activated.
Stop
• Press OK for approx. 1 second. The temperature
control is deactivated.
• Switch off the unit with the circulator's mains
power switch.
In combination with the cold bath base units 600F,
601F, 900F, 1000F, the flashing control indicator
"Cooling" after temperature regulation is
complete indicates that a valve is still open. Only
turn off the main power switch after this indicator
has stopped flashing.
The button is used to set device parameters in the dialog.
Here are the menus:
Int Selection of the interface
Pu Pump capacity
tset Adjustment of the timer
SAFE Safety adjustments
Atc Temperature sensor adjustment
conF Device configuration
data Data processing
The setting options of the menus are described in the chapters.
9.3 Excess temperature and low level safety devices
The safety devices are not affected by the control circuit. When they
trip all actors are permanently shut down and a start of the
temperature control is no longer possible.
The alarm is displayed optically and acoustics with a continuous
signal tone and the reason for the alarm is shown on the display as
a number.