Preparations for operating the device
15
7.1 Installation of the circulator
Danger of scalding due to leaks from the baths
The JULABO plastic baths are not solvent-resistant. JUALBO plastic
bath tanks are for water at a working temperature range from
+20°C to +100°C.
Do not contaminate the bath fluid with solvents.
Things to keep in mind during the installation process:
Risk of tipping due to improper transportation
Crushing, damage to the unit
• Use PPE (safety gloves, safety shoes, safety goggles).
• Carry the unit with 2 persons (see the Technical Data for the
weight).
• Transport the unit carefully on firm, level ground. The interior
of the unit can also be damaged by impacts or if it is dropped.
• Do not loiter under the unit during transportation and
operation.
• The installation site should be a sufficiently large room to
ensure that it does not become too hot due to the heat
emission.
• The surface for the device should be flat and made of non-
flammable material.
• A specific room size is prescribed for refrigerated circulators.
• At high temperatures, position the unit under an exhaust hood
as much as possible due to potential vapors from the thermal
bath fluid.
• Observe the safety sticker - do not remove!
8 Preparations for operating the device
8.1 Securing the immersion circulator
Danger of electric shock.
Switch off the device and unplug it from the power
supply
Carefully secure the immersion circulator on the bath vessel. Poorly
installed circulators can fall into the bath tank. If this happened have
the unit checked by a service technician prior to re-use.