Commissioning
33
9.8 Device configuration
5x
conF
In this menu is adjustable:
- Auto Auto start On / Off
- hLiN Supply voltage On / Off
- cooL Cooling mode On / Off / Auto
- init Factory Init
Switch Autostart on and off
The circulator has been configured and delivered by JULABO
according to N.A.M.U.R. recommendations. This means for the start
mode that the unit must enter a safe operating state after a power
failure (non-automatic start mode). This safe operating state is
indicated by “OFF” or “R OFF” on the display. A complete shutdown
of the main functional elements such as the heater and circulating
pump is effected simultaneously.
The values set on the circulator remain stored, and the unit is
returned to operation by pressing the
key (in manual control
mode).
In remote control mode, the values need to be resent by the PC via
the interface.
Should such a safety standard not be required, the AUTOSTART
function (automatic start mode) may be activated, thus allowing the
unit to be started directly by pressing the mains power switch or
using a timer.
The AUTOSTART function allows the device to be started as soon as
the main switch is pressure, which in turn allows you to use a timer.
Uncontrolled device start
If circulators are started using "AUTOSTART", ensure that even if it
is started accidentally, for example after a power outage, it does not
pose a danger to personnel or equipment.
• Ensure that the circulator's safety equipment is set correctly.
Auto
1. Select submenu Auto.
2. Press .
3. Select On or OFF with or ,
4. Confirm with
(The display flashes briefly).