3 Description
• These circulators are operated via the splash-proof keypad. The
microprocessor technology allows the setpoint to be set,
displayed and saved using the temperature display LED.
• The PID temperature control automatically adjusts the heat
supply to the requirements in the bath.
• ATC - Absolute Temperature Calibration (3-point calibration)
• The pump function can be switched from internal to external
circulation by a simple switchover button.
• Safety equipment to IEC 61010-2-010:
Excess temperature protector is a safety device which is
independent of the control circuit whose value is set
using a tool (screwdriver).
A float switch acts as the low level safety device. If these safety
devices trip, the heater and circulating pump are
completely shut down.
• Interfaces:
- CAN bus to communicate with the refrigeration machine
- RS232 interface for remote control.
- USB connector to act as host to update and read data. USB
device for remote control.
It is important to follow these safety instructions to prevent personal
injury and property damage. These instructions apply in addition to
the safety instructions at your workstation.
It is essential that you read the user information before
starting the device.