EasyManua.ls Logo

Julabo CORIO CD-900F - Cleaning; 17 Storage

Julabo CORIO CD-900F
68 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Storage
55
16.2 Cleaning
Use low surface tension water (for example soap suds) to clean the
bath and the functional parts of the circulator which are immersed in
it. Clean the exterior device with a cloth and low surface tension
water.
The circulator is designed for continuous use in normal conditions.
No regular maintenance work is required.
The bath tank should only be filled with suitable bath fluid. In the
event of contamination, the bath fluid must be replaced from time to
time.
Cleaning open bath tanks
Leaking bath tanks due to unsuitable cleaning products.
These bath tanks are not resistant to solvents and pure
alcohol. Incorrect cleaning products will make the surface of
the bath go cloudy and will dissolve the adhesive. Plastic
baths will therefore start to leak.
Clean bath tanks with wet products - never rub them dry. The
cloths or sponges you use should not be contaminated (with
scouring particles or dust).
A good, essentially smear-free cleaning effect can be
achieved using a microfiber cloth moistened with water.
Use warm water with a few drops of washing-up liquid and a
soft cloth to clean the bath.
If it has heavier soiling, particularly if it is greasy, benzol-free
pure benzene (washing benzene, light benzene) can be used
for cleaning.
17 Storage
Units which are not to be reused must be stored in a dry, place,
protected from dust and frost, after cleaning. The system
components must be fully emptied and carefully dried, for example
using compressed air. Seal the connectors.