Commissioning
34
Power supply
The circulator is supplied with power via the refrigeration base unit
(fig. a). However, circulator and refrigeration base unit can be
supplied separately (fig. b). The heating capacity limit of the
circulator is deactivated. Thus, more capacity is available for heating
processes with separate fuse protection.
The circulator is powered by the basic cooling unit. However, there is
the possibility of a separate supply of circulator and basic cooling
unit.
hLiN
1. Select submenu hLiN.
2. Press
.
3. Select
yEs (factory setting) or nowith or .
4. Confirm with
(Display flashes four times).
The switchover process is shown briefly on the temperature display.
Reset parameter (Factory init)
With the factory init all settings are reset to the values of the
delivery. The ATC calibration values are deleted.
init
1. Select submenu init.
2. Press
.
3. Select
yes or noFwith or .
4. Confirm with
(the display flashes briefly).
The switchover process is shown briefly on the temperature display.