Error messages / Possible causes of faults
44
CFG-
Error 3
Configuration file corrupt. Contact JULABO Service Department..
ProG-
Error 1
Invalid data detected. Please repeat process. If it does appear again,
change the USB stick or contact JULABO Service.
ProG-
Error 2
Starting the programming failed. Please repeat process. If it does
appear again, change the USB stick or contact JULABO Service.
ProG-
Error 3
Errors occurred during deletion. Please repeat process. If it does
appear again, change the USB stick or contact JULABO Service.
ProG-
Error 4
Error while writing or verifying occurred. Please repeat process. If it
does appear again, change the USB stick or contact JULABO Service.
Error 5
Update aborted. Please repeat process. If it does appear again,
change the USB stick or contact JULABO Service.
To cancel the alarm state
1. Switch off the device at the main switch.
2. Eliminate the cause of the alarm or wait for approx. 4
seconds, depending on error type.
3. Switch on the device again at the mains power switch.
4. 4. If the error occurs again, a remote diagnosis must be
made by the JULABO Service. To do this, provide a black
box file (see chapter 10.1).
Faults which are not displayed:
Circulating pump motor overload protection.
The circulating pump motor is protected from overloads. If the motor
is blocked it switches off, the warning E1305 is generated. To
remedy this, stop the temperature control, remove the cause of the
blockage and restart tempering.
If necessary the unit should be inspected by a JULABO service
technician.
JULABO Technical Service
Phone: +49 07823 5166
Fax: +49 07823 5199
Email: service.de@julabo.com