No counting
Display reading in case of blocking of flow
After a blocking of the installation, the flow display in MULTICAL® 803 will drop to 0 l/h within a few seconds when a
flow sensor with quick pulses such as ULTRAFLOW® is used.
When MULTICAL® 803 is connected to flow sensors with slow pulses, e.g. reed switch, the flow display first shows a
decreasing flow several minutes after a blocking. MULTICAL® 803 automatically sets the flow display to 0 l/h after 60
minutes without pulses. For flow sensors with slow pulses, the flow display generally responds slower and is less
suitable for displaying low flows than when using flow sensors with quick pulses.