Kamstrup A/S · Technical description · 5512-2360_A1_GB_02.2019
Diagram 1: Total typical accuracy of MULTICAL® 803, sensor pair and ULTRAFLOW® compared to EN 1434-1.
2.4 Electrical data
Calculator: E
c
(0.15 + 2/
) % Sensor pair: E
t
(0.4 + 4/
) %
LCD with white LED (32 x 63 mm) - 7 or 8 digits with 10 mm digit height
999.9999 - 9999.999 – 99999.99 – 999999.9 – 9999999
9999.9999 - 99999.999 – 999999.99 – 9999999.9 – 99999999
programmable
Logging intervals: From one minute to one year
Logger content: All registers can be selected
Standard logger profile: 20 years, 36 months, 460 days, 1400 hours
280 info codes can be read out via LogView, the latest 50 info codes can be read on the
calculator's display
(with backup battery)
Clock, calendar, leap year compensation, target date
Programmable under country code
The function can be disabled so that “technical normal time” is used.
Without external adjustment: Deviation less than 15 min./year from legal time
With external adjustment Deviation less than 7 s from legal time
KMP protocol with CRC16 used for optical communication
as well as modules.
sensors
10
W RMS
-6,0
-4,0
-2,0
0,0
2,0
4,0
6,0
0,01 0,10 1,00 10,00
Tol. [%]
q [m³/h]
MULTICAL
®
803 and ULTRAFLOW® qp 1.5 m³/h @
30K
Ec+Et+Ef (EN) Ec+Et+Ef (Typ)