SCAN EXAM B-1
B. Installation and setup
Appendix B. Installation and setup
Installation and setup must only be
carried out by service personnel trained
and approved by the manufacturer of
the unit.
B.1 System installation
B.1.1 Positioning the unit
Position the unit on a stable fl at surface so
that vibrations will not degrade the image
quality.
The unit can also be attached to a wall, under
or on a shelf using the optional scanner
mounting kit (pt. no. 204138).
Do not position the unit in direct sunlight or
near bright light. Sunlight or bright light must
not be allowed to shine directly on the unit
door into which the IPs are inserted.
The unit must not be positioned so that it
touching other equipment. The unit must
not be placed on top of or under other
equipment.
The unit can be positioned within the
environment in which the patient is examined
and treated (patient environment).
B.1.2 Positioning the PC(s) (not supplied)
The PC(s) connected to the unit should not
be used in the patient environment.
The minimum horizontal distance between
the patient and the PC(s) is 1.5 m (4.5 ft).
The minimum vertical distance between the
patient and the PC(s) is 2.5 m (6.5 ft).