A-8 Specifications Model 2790 SourceMeter
®
Switch System User’s Manual
Ohms specification calculations (7751/7752/7753)
Ohms specifications for the 7751/7752/7753 module provide the ohms accuracy and tem-
perature coefficient data for specific I-source and V-source output values. The following
formulas can be used to determine the ohms specification for any valid I-source or V-source
output setting not published in the specification sheet.
Ohms accuracy and temperature coefficient are specified as percent of reading plus an
ohms offset.
7751/7752/7753 I-source
Low ohms specifications using the I-source of the 7751/7752/7753 are provided for the
following current output settings: 50mA, 20mA, 10mA, and 1mA (dry circuit ohms). Use
the following formulas for other I-source settings:
NOTE The following formulas are only valid when using 1VDC range of the
Model 2790 DMM. When using the SIxOhms math calculation, the 1V range is
selected automatically.
Ohms accuracy
% of reading = (0.07 + 1/I
SOURCE
) %
Ohms offset = (0.03/I
SOURCE
)
where; I
SOURCE
is the current output setting in mA
Example – Determine ohms accuracy for an I-source setting of 30mA:
% of reading = (0.07 + 1/30) %
= (0.07 + 0.03) %
= 0.10%
Ohms offset = (0.03/30)
= 1m
Ohms accuracy (30mA) = 0.10% + 1m