2-8 Model 2790 SourceMeter
®
Switch System User’s Manual
Remote programming
As shown in Table 2-1, the amplitude command is used to set the outputs of the sources.
Notice that the channel number of the V-source (28) and I-source (27) must be included in
the command string.
Table 2-1
Commands to set amplitude for V-source and I-source
Commands Description Default
SOURce:VOLTage[:LEVel][:IMMediate]
[:AMPLitude] <NRf>, <clist> Set V-source output; 50 to 500 (V).
<clist> = (@128) Slot 1
= (@228) Slot 2
50V
[:AMPLitude]? <clist> Query V-source output level.
SOURce:CURRent[:LEVel][:IMMediate]
[:AMPLitude] <NRf>, <clist> Set I-source output; 0 to 50 (mA).
<clist> = (@127) Slot 1
= (@227) Slot 2
0.001A
[:AMPLitude]? <clist> Query I-source output level.
Note: Selecting an ohms calculation enables source tracking. Selecting any other math function disables
source tracking.