EasyManua.ls Logo

Keithley 2790 - Figure 5-4 Test Connections - Shunt Bar and Bridgewire Bar Testing - Dual Stage Inflator

Keithley 2790
130 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Model 2790 SourceMeter
®
Switch System User’s Manual 5-7
Dual stage inflator test connections:
Figure 5-4 — Use this connection scheme if using a Keithley 7752 module. It can
also be used for the 7751 or 7753 module if not performing the HIPOT test.
Figure 5-5 — Use this connection scheme if using a Keithley 7751 or 7753 mod-
ule. It includes the connections for the HIPOT test.
Figure 5-4
Test connections – shunt bar and bridgewire bar testing – dual stage inflator
SRC HI
SEN HI
SRC LO
SEN LO
Ch 1
Ch 2
Bank 1
J101
SRC HI
SEN HI
SRC LO
SEN LO
Ch 4
Ch 5
Bank 2
J102
A
B
Shunt
Bars
Dual Stage Inflator
B) Bridgewire Testing
A) Shunt Bar Testing
A
B
Some wires are shown in bold
to make it easier to follow the
signal paths.
NOTE
Keithley 7751, 7752, or 7753

Table of Contents

Related product manuals