2-18 Model 2790 SourceMeter
®
Switch System User’s Manual
Manual cable discharge
The discharge circuit can also be activated manually at any time by closing channel 20 of
the 7751/7753 module. Typically, manual discharge is used when a discharge time >5msec
is desired.
Figure 2-6
Cable discharge circuit
100k
CSO
~1nF
C
SE
~1nF
100k
Source
Sense
HI
LO
Source
Sense
Inflator
Under
Test
= Closed channel switches.