Example 2
This example is for a partially loaded Model 3706 with Model 3723 cards (all 1 pole mode).
Slot #
Card
Channel relays
closed
Backplane relays
closed
SLOT 1
3723
10
7
1
SLOT 2
3723
107
1
SLOT 3
EMPTY
0
0
SLOT 4
3723
107
1
SLOT 5
3723
107
1
SLOT 6
EMPTY
0
0
This produces the following power consumption:
Slot 1 Power Consumed =
700
+
107 x 50
+
1 x 100
=
6150
Slot 2 Power Consumed =
700
+
107 x 50
+
1 x 100
=
615
0
Slot 3 Power Consumed =
0
+
0
+
0
=
0
Slot 4 Power Consumed =
700
+
107 x 50
+
1 x 100
=
6150
Slot 5 Power Consumed =
700
+
107 x 50
+
1 x 100
=
6150
Slot 6 Power Consumed =
0
+
0
+
0
=
0
Totals for each bank can then be calculated as follows:
Slot
1
Slot 2
Slot 3
Total
Bank #1 Power Consumed =
6150
+
6150
+
0
=
12300
Slot 4
Slot 5
Slot 6
Total
Bank #2 Power Consumed =
6150
+
6150
+
0
=
12300
Result: Since each bank has not exceeded maximum power, this is OK.

Table of Contents

Other manuals for Keithley 3730

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 3730 and is the answer not in the manual?

Keithley 3730 Specifications

General IconGeneral
BrandKeithley
Model3730
CategoryMultimeter
LanguageEnglish

Related product manuals