Table 7-4. Amps Calibration
617 Calibration R;;ei;or Calibrator 617 Display IEEE-488
Range current
l
ValUe Value
Bus Command
ZOOpA
I*
190.00 pA ~ -__
l *
190.00 pA
A190E-12X
20 nA 19.00 nA lO.OOMR 0.19OOOV , 19.000 nA ' A19E-9X
20j4A
19.00 pA l.OOOMQ 19.OOOOV 19.000 /LA A19E-6X
20mA 19.000mA 1000.0 Q 19.OOOOV 19.000mA A19E-3X
SHIELDED FIXTURE SHIELDED FIXTURE
,SEE FIGURE 7-l) ,SEE FIGURE 7-l)
-L -L
r----
GROUND LINK IN PLACE
ITE: LEAVE GREEN DISCONNECTED
I
BLACK
LINK REMOVED
Figure 7-5. Connections For Amps Calibration
(20nA. 20pA and 20mA Ranges1
Figure 7-4. Connections For Amps Calibration
(20pA)
7-7