EasyManua.ls Logo

Keithley DMM6500

Keithley DMM6500
99 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Model DMM6500 -Digit Multimeter with Scanning Calibration and Adjustment Manual Section 2: Performance verification
DMM6500-905-01 Rev. A/July 2018 2-23
Four-wire resistance verification with offset compensation off
The values and limits in the following tables are for example only. You must calculate test limits
based on the actual resistance values output by your calibrator or resistance source (see Example
reading limit calculation (on page 2-4)).
For 10 MΩ verification, the Sense HI cable is optional. Measurement is with Input HI and LO and
Sense LO only.
Range
(Ω)
Nominal calibrator
values
(Ω)
Actual calibrator
value
(Ω)
Lower limit
(Ω)
Upper limit
(Ω)
1.0E+05
1.00E+05
1.0000060E+05
9.99921E+04
1.000091E+05
1.0E+06
0.00E+00
n/a
6.00E+00
6.00E+00
1.00E+06
9.99851E+05
1.000063E+06
1.000063E+06
1.0E+07
0.00E+00
n/a
1.00E+02
1.00E+02
1.00E+07
9.9989960E+06
9.99440E+06
1.000360E+07
Verify 2-wire resistance accuracy
To verify the 2-wire resistance function 100 MΩ range, you will:
Use shielded, Teflon-insulated or equivalent cables in a 2-wire configuration.
Apply accurate resistance from the calibrator to the terminals on the front panel of the DMM6500.
Verify that the displayed readings are within specified limits.
Verify resistance 100 MΩ range
To verify the 100 MΩ range:
1. Connect the DMM6500 INPUT to the calibrator as shown in the following figure.
Figure 11: Connections for 100 MΩ verification

Table of Contents

Other manuals for Keithley DMM6500

Related product manuals