Section 3: Adjustment Model DMM6500 -Digit Multimeter with Scanning Calibration and Adjustment Manual
3-18 DMM6500-905-01 Rev. A/July 2018
Command parameters
Calibrator
function
Calibrator
range
Calibrator
value
Calibrator
stimulus
value*
Calibrator
stimulus
value
1 mA range (+) full scale nominal
cal_DCI_1mA_fs_pos
DC current
0.0022
0.001
8508A
Reading
1 mA
cal_DCI_DIGI_1mA_fs_pos
DC current
0.0022
0.001
8508A
Reading
1 mA
1 mA range (-) full scale nominal
cal_DCI_1mA_fs_neg
DC current
0.0022
-0.001
8508A
Reading
1 mA
cal_DCI_DIGI_1mA_fs_neg
DC current
0.0022
-0.001
8508A
Reading
1 mA
10 mA range (+) full scale nominal
cal_DCI_10mA_fs_pos
DC current
0.022
0.01
8508A
Reading
10 mA
cal_DCI_DIGI_10mA_fs_pos
DC current
0.022
0.01
8508A
Reading
10 mA
10 mA range (-) full scale nominal
cal_DCI_10mA_fs_neg
DC current
0.022
-0.01
8508A
Reading
10 mA
cal_DCI_DIGI_10mA_fs_neg
DC current
0.022
-0.01
8508A
Reading
10 mA
*Characterized step measured by the Fluke 8508A
Command parameters
Calibrator
function
Calibrator
range
Calibrator
value
Calibrator
stimulus
value
Reference
DMM DC
current range
cal_DCI_100mA_fs_pos
DC current
0.22
0.1
n/a
20 A
cal_DCI_DIGI_100mA_fs_pos
DC current
0.22
0.1
n/a
20 A
cal_DCI_100mA_fs_neg
DC current
0.22
-0.1
n/a
20 A
cal_DCI_DIGI_100mA_fs_neg
DC current
0.22
-0.1
n/a
20 A
cal_TS5
DC current
n/a
n/a
n/a
20 A
cal_DCI_1A_fs_pos
DC current
2.2
1
n/a
20 A
cal_DCI_DIGI_1A_fs_pos
DC current
2.2
1
n/a
20 A
cal_DCI_1A_fs_neg
DC current
2.2
-1
n/a
20 A
cal_DCI_DIGI_1A_fs_neg
DC current
2.2
-1
n/a
20 A
cal_DCI_DIGI_3A_fs_pos
DC current
2.2
2
n/a
20 A
cal_DCI_DIGI_3A_fs_neg
DC current
2.2
-2
n/a
20 A
DC current 10 A range adjustment
The following procedure provides instructions for completing a rear-terminal DC current adjustment
for the 10 A range. The following section provides a command parameter table to complete the
adjustment.

Table of Contents

Other manuals for Keithley DMM6500

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley DMM6500 and is the answer not in the manual?

Keithley DMM6500 Specifications

General IconGeneral
Resolution6.5 digits
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
DC Current Range10 µA to 10 A
AC Current Range10 µA to 10 A
Temperature Range-200 °C to 1820 °C
Display5-inch capacitive touch screen
Resistance Range100 MΩ
Frequency Range3 Hz
Sampling Rate1 MS/s
ConnectivityUSB, LAN, GPIB (optional), TSP-Link
Memory7 million readings