In this section:
Introduction .............................................................................. 3-1
Environmental conditions ......................................................... 3-1
Warmup period ......................................................................... 3-2
Adjustment overview ................................................................ 3-2
Recommended test equipment ................................................ 3-3
General adjustment considerations .......................................... 3-3
Initial instrument setup ............................................................. 3-4
Remote calibration adjustment procedures .............................. 3-6
Enable temperature correction ............................................... 3-37
Save calibration and set the adjustment dates ....................... 3-38
Setting time, adjustment, and verification dates ..................... 3-38
Adjustment command timing and error checking ................... 3-39
Handling events ..................................................................... 3-39
Introduction
Use the procedures in this section to adjust the DMM6500 calibration.
DMM6500 performance is specified for a period of 90 days, 1 year, or 2 years from adjustment.
Adjustment should be performed at one of these intervals, depending on your specification
requirements.
Performance is specified relative to calibration adjustment temperature (T
cal
). Keithley factory
adjustment is performed at 23 °C ± 1 °C.
The information in this section is intended for qualified service personnel only. Do not
attempt these procedures unless you are qualified to do so. Some of these procedures may
expose you to hazardous voltages, which could cause personal injury or death if contacted.
Use appropriate safety precautions when working with hazardous voltages.
Environmental conditions
To make sure accurate results, the environment must meet the following conditions.
Section 3
Adjustment

Table of Contents

Other manuals for Keithley DMM6500

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley DMM6500 and is the answer not in the manual?

Keithley DMM6500 Specifications

General IconGeneral
Resolution6.5 digits
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
DC Current Range10 µA to 10 A
AC Current Range10 µA to 10 A
Temperature Range-200 °C to 1820 °C
Display5-inch capacitive touch screen
Resistance Range100 MΩ
Frequency Range3 Hz
Sampling Rate1 MS/s
ConnectivityUSB, LAN, GPIB (optional), TSP-Link
Memory7 million readings