EasyManua.ls Logo

Keithley DMM6500

Keithley DMM6500
99 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Section 2: Performance verification Model DMM6500 -Digit Multimeter with Scanning Calibration and Adjustment Manual
2-36 DMM6500-905-01 Rev. A/July 2018
Figure 18: Front panel 4-wire shorting plug orientation
8. Verify that the 1 Ω range is within specification (see the following table).
9. Repeat verification for the 10 Ω to 100 kΩ ranges.
Verify 4-wire resistance zero values
Range (Ω)
Lower limit
Upper limit
1
2.00E-04
2.00E-04
10
2.00E-04
2.00E-04
100
2.00E-03
2.00E-03
1 k
6.00E-03
6.00E-03
10 k
6.00E-02
6.00E-02
100 k
1.00E+00
1.00E+00
Verify DC voltage zero values using the 4-wire short
To verify DC voltage zero values:
1. Leave the short connected as described in Verify resistance zero values using a 4-wire short (on
page 2-35).
2. Press the FUNCTION key.
3. Select the DC Voltage function.
4. Press the HOME key.
5. Set the range to 1000 V.
DC voltage verification is done in descending range order, starting with the 1000 V range and
finishing on the 100 mV range.
6. Verify that the 1000 V range zero is within specification. See the table below.
7. Verify that the 100 V to 100 mV range zero is within specification.

Table of Contents

Other manuals for Keithley DMM6500

Related product manuals