EasyManua.ls Logo

kENDALL KANGAROO ePump - Intermittent Mode (Adjust Bolus); Intermittent Mode Feeding

kENDALL KANGAROO ePump
42 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
18
KANGA
R
OO
ROO R
e
P
um
p
English English
Intermittent Mode (Adjust Bolus)
Intermittent Mode Feeding
For
INTERMITTENT
feeding mode, the
ADJUST BOLUS
option is available in the opening menus (Figure 7C for
a Feed-Only set or Figure 7D for a Feed & Flush set). Select this option to program the feeding parameters from the
ADJUST BOLUS
screen, Figure 12A.
ADJUST BOLUS screen, Figure 12A. ADJUST BOLUS
Press
4
BOLUS RATE
in Figure 12A to define the rate of delivery.
Press the
4
buttons in Figure 12B to program the delivery rate from 1 to 400 in increments of 1 ml/hr.
Press
4
ENTER
(top button) to exit the menu.
Press
4
NUMBER OF BOLUSES
in Figure 12A to define the total number of boluses to be delivered.
Press the
4
buttons in Figure 12C to program the number of boluses, up to a maximum of 99.
Press
4
ENTER
(top button) to exit the menu. The number of boluses can be set as zero, in which case the pump will
run until the supply is exhausted.
Press
4
VOLUME PER BOLUS
in Figure 12A to define the Volume to be delivered in each bolus.
Press the
4
buttons in Figure 12D to program the volume per bolus from 1 to 1000 ml in increments of 1 ml.
Press
4
ENTER
(top button) to exit the menu.
Press
4
BOLUS INTERVAL
in Figure 12A to define the time interval between starts of bolus deliveries.
Press the
4
buttons in Figure 12E to program the time interval between starts of bolus deliveries from 1 to 24 hours in
increments of 1 hour. Press
4
ENTER
(top button) to exit the menu.
The pump will automatically limit bolus settings which
exceed pump capabilities
to deliver. For example, a bolus
delivery volume of 200 mL cannot be made once per hour if a rate of 150 mL/hr is already programmed. In this case,
the maximum volume allowed would be 150 mL.
NOTE:
If the
NUMBER OF BOLUSES
is reprogrammed during a feeding (from HOLD mode), the pump will restart
the bolus feeding and will deliver all the boluses as defined by the new setting.
© 2005 Tyco Healthcare Group LP

Table of Contents