Intermittent Mode (Adjust Bolus)
Intermittent Mode Feeding
option is available in the opening menus (Figure 7C for
a Feed-Only set or Figure 7D for a Feed & Flush set). Select this option to program the feeding parameters from the
ADJUST BOLUS screen, Figure 12A. ADJUST BOLUS
in Figure 12A to define the rate of delivery.
buttons in Figure 12B to program the delivery rate from 1 to 400 in increments of 1 ml/hr.
(top button) to exit the menu.
in Figure 12A to define the total number of boluses to be delivered.
buttons in Figure 12C to program the number of boluses, up to a maximum of 99.
(top button) to exit the menu. The number of boluses can be set as zero, in which case the pump will
run until the supply is exhausted.
in Figure 12A to define the Volume to be delivered in each bolus.
buttons in Figure 12D to program the volume per bolus from 1 to 1000 ml in increments of 1 ml.
(top button) to exit the menu.
in Figure 12A to define the time interval between starts of bolus deliveries.
buttons in Figure 12E to program the time interval between starts of bolus deliveries from 1 to 24 hours in
increments of 1 hour. Press
(top button) to exit the menu.
The pump will automatically limit bolus settings which
to deliver. For example, a bolus
delivery volume of 200 mL cannot be made once per hour if a rate of 150 mL/hr is already programmed. In this case,
the maximum volume allowed would be 150 mL.
is reprogrammed during a feeding (from HOLD mode), the pump will restart
the bolus feeding and will deliver all the boluses as defined by the new setting.