Section X — Specifications and Symbols
KANGAROO ePump MISTIC Feed-Only Set or Feed & Flush Set
Feeding Formula Delivery Rate
1-400 ml/hr in 1 ml increments
1-3000 ml in 1 ml increments
1-24 hours in 1-hour increments
Medical Electrical Equipment
KANGAROO ePump™ Enteral Feeding Pump
(1) Classified with respect to electrical shock, fire, and mechanical hazards in
accordance with UL60601-1.
(2) Classified with respect to electrical shock, fire, mechanical and other specified
hazards in accordance with CAN/CSA C22.2 No. 601.1
The pump status LED array on the upper right of the pump gives a quick visual indication of the pump status, especially in
A green light status indicates normal pump operation.
A yellow light status indicates an informational or warning situation for a low battery condition, the pump in
mode, or one of the following errors:
A red light status indicates a critical pump error. The screen will show one of the following error messages:
© 2005 Tyco Healthcare Group LP