34
KANGA
R
OO
ROO R
e
P
um
p
English English
Section X — Specifications and Symbols
Specifications
UL Listing
Type Infusion Device
Volumetric
Pumping Mechanism
Rotary Peristaltic
Pump Sets
KANGAROO ePump MISTIC Feed-Only Set or Feed & Flush Set
Feeding Formula Delivery Rate
1-400 ml/hr in 1 ml increments
Feeding Formula VTBD
1-3000 ml in 1 ml increments
B
olus Volume
1-1000 ml in 1 ml
increments
Number of Boluses
1-99
Bolus Interval
1-24 hours in 1-hour increments
Medical Electrical Equipment
KANGAROO ePump™ Enteral Feeding Pump
(1) Classified with respect to electrical shock, fire, and mechanical hazards in
accordance with UL60601-1.
(2) Classified with respect to electrical shock, fire, mechanical and other specified
hazards in accordance with CAN/CSA C22.2 No. 601.1
LED Array
The pump status LED array on the upper right of the pump gives a quick visual indication of the pump status, especially in
darkened rooms.
A green light status indicates normal pump operation.
A yellow light status indicates an informational or warning situation for a low battery condition, the pump in
HOLDING
mode, or one of the following errors:
HOLD ERROR
,
FEED ERROR
,
FLOW ERROR
, and
FLUSH ERRO
R
.
A red light status indicates a critical pump error. The screen will show one of the following error messages:
SYSTEM
ERROR
,
PUMP SET
DISLODGED
, or
ROTOR ERROR
.
© 2005 Tyco Healthcare Group LP