Enteral Feeding
Pump
English English
Flushing Solution Dose Range
10-500 ml in 1 ml increments
Flushing Solution Interval Range
1-24 hr in 1 hr increments
Accuracy
±
10% or 0.5 ml/hr, whichever is larger, with bag at 46 cm (18”) above pump, at room temperature
2
C
±
2° C (
72°
F
±
3° F),
using a new pump set with less than the recommended 24 hours of maximum usage.
Occlusion Pressure
15 psi (103 kPa) Maximum
Dimensions
Height: 16.8 cm (6.6”) Width: 16.3 cm (6.4”) Depth: 11.7 cm (4.6”)
Weight
1.1 kg (2.4 lbs), 1.2 kg (2.7 lbs) with pole clamp
Material
Housing: ABS/Polycarbonate blend
Pump Set Door: Polyester/ Polycarbonate blend
Power
Use AC adapter for wall outlet usage. Pump operates on 9V DC, 1.5 Amps.
Battery
New, fully charged NiMH Battery Pack delivers ≥ 15 hrs at 100 ml/hr feed rate. Approximately 15 minutes prior
to complete battery discharge, a low battery alarm will occur (see subsection “Battery Low” in Section IX). When
complete discharge occurs, the pump will automatically turn itself off.
The battery will charge continuously whenever the pump is plugged into a wall outlet. Eight hours of charging is
required to fully recharge the battery pack.
Alarms
System Error
Flow Error
Pump Set Dislodged
Error
Hold Error
Rotor Error
Feed Error
Flush Error
Battery Low
Alarm Volume
Minimum of 70 dBA at 1 meter in maximum volume orientation
© 2005 Tyco Healthcare Group LP