to return to one of the opening menus shown in Figure 7, thus allowing any setting to be
modified. If the Pump Set has been removed after going into
, the opening menu will show
LOAD A SET, as seen LOAD A SET
Resume After Specified Time Interval
to immediately input a time in minutes in which to resume operation. Each press of
will increment the resume time by 5-minute increments, up to 30 minutes. After “30” is
reached, each additional button press will increment in larger time steps, up to a maximum of 240 minutes. Another
button push after 240 will cycle the “resume in” time back to zero. No other input is needed to start the countdown, as
the time immediately counts down after the incremental value is reached.
to immediately restart pump operation. This option will not be available if the Pump Set is removed during
HISTORY to access the HISTORY
screen. Use this screen to review feeding and flushing history. See subsection
ORY screen. Use this screen to review feeding and flushing history. See subsection ORY
“More Options” below for further information.
mode, the pump can be diverted to immediately perform a flush (flush on demand) by
RUNNING mode, the pump can be diverted to immediately perform a flush (flush on demand) by RUNNING
RUNNING menu. See Figure 14.RUNNING
will prompt for the volume to be immediately flushed, and will show the previously programmed flush volume as
default. The screen will appear as the standard Flush input screen as shown in Figure 11B (or 13B). Setting the
flushing volume is temporary and will not change the main/periodic flushing volume previously programmed
is pressed by accident, the screen will return to the
screen after 10 seconds, if there is no
RUNNING screen after 10 seconds, if there is no RUNNING
input for the flush volume.
© 2005 Tyco Healthcare Group LP