General Cleaning Directions
Cleaning of KANGAROO ePump
pumps must be performed as follows:
CAUTION: Disconnect pump from AC power source before cleaning. After cleaning, do not connect to
AC power source until pump and power cord are thoroughly dry.
• A mild detergent should be used for general cleaning. If necessary, the pump may be cleaned with a 10:1 water and
hypochlorite mixture, however, repeated cleaning with this solution can damage the plastic pump case. Isopropyl
alcohol, applied with a damp (not wet) cotton swab, may be used for cleaning difficult-to-reach areas, however, it
should be used sparingly because repeated cleaning may damage the case.
• DO NOT USE strong cleaners such as Spray-Nine™*, Phisohex™*, Hibiclens™*, or Vesta-Syde™* because damage to
the pump case housing can result.
Directions for Cleaning Pump Housing
• Refer to General Cleaning Directions before starting.
• Clean outside surface with a damp (not wet) cloth or sponge and keep pump in upright position as much as
• Avoid excess moisture near pole clamp area.
• Do not allow cleaning solution to enter the vertical vents in the housing, underneath the handle.
Directions for Cleaning Pump Power Cord
• Refer to General Cleaning Directions before starting.
• Unless soiling is observed, the power cord should not be cleaned.
• If cleaning of the power cord is necessary, unplug from outlet and wipe the exterior surfaces of the wall plug with a
cloth dampened with isopropyl alcohol.
Avoid exposing power cord to excess moisture, as this can lead to an electrical shock or
Directions for Cleaning Rotor Assembly
Refer to General Cleaning Directions before starting.
• Open the blue door enclosing the Pump Set loading area.
Loosen rotor set screw with 5/64” (2mm) allen wrench and gently pull rotor off shaft. After removing rotor, avo
getting any moisture in the rot
roughly with warm soapy water, or isopropyl alcohol if necessary.
Be sure all parts of rotor are completely dry before putting it back onto shaft.
To replace rotor, align set screw on hub of rotor with the flattened section of the
Push rotor into place and tighten set screw (Do not over tighten)