CFS/CCS-BA-e-1220 27
7.2 Adjustment reference balance CFS 50K-3 models
Switch-on balance and during the selftest press .
Use the numeric keys to enter password:
Enter sandard password „0000“
Confirm input by .
When using as counting system, the bulk material
balance as well as the reference balance must be
adjusted. The adjustment process must be carried out
on both balances.
With select reference balance („LoCAL“). The
appeared [] displays the current balance.
Acknowledge with .
If necessary, at balance zero display using select
the weighing unit [kg or lb], which shall be used for
adjustment. The appeared [] displays the current
weighing unit.
Acknowledge with .
Ensure that there are no objects on the weighing pan.
After successful standstill control press the button.
Place the required adjusting weight carefully in the
centre of the weighing plate and press to confirm.
After the adjustment the balance will carry out a self-test.
Remove adjusting weight during selftest, balance will
return into weighing mode automatically.
In case of an adjustment error or incorrect adjusting
weight the display will show an error message
( ), repeat adjustment process.