9.4 Weighing with tare
A tare value can be entered for the reference as well as for the bulk material
balance. Before setting a tare values select active balance, s. chap. 9.3.
9.4.1 Taring
Deposit weighing vessel. After successful stop check press the TARE button.
The zero display and the symbol [] above NET will appear.
The weight of the container is now internally saved.
Weigh the material, the net weight will be indicated.
After removing the weighing container, the weight of the weighing container
appears as negative display.
To delete the tare value, remove load from weighing pan and press the TARE-
key.
The tare procedure can be repeated as many times as necessary, for example
with initial weighing of several components for a mix (add-on weighing). The
limit is reached when the total weighing range capacity is full.
9.4.2 Numeric entering of tare weight
Unload and reset to zero the balance
Input known tare weight numerically with decimal point and press key TARE.
The entered weight will be stored as tare weight and displayed with negative
sign. The [] symbol over NET appears.
Put the filled weighing container on the balance, the net weight will be
displayed.
The tare value remains stored until it will be deleted by TARE.
The tare value will be rounded up according to the readability of the balance,
e.g. at a balance 60 kg max/5 g readability the input value of 103 g will be
displayed as -105 g.

Table of Contents

Other manuals for KERN CFS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the KERN CFS and is the answer not in the manual?

KERN CFS Specifications

General IconGeneral
Plate materialstainless steel
Readability0.01 g
Reproducibility0.01g
CalibrationExternal calibration
DisplayLCD
Weighing unitsg, lb, oz
Operating temperature10 °C - 40 °C
Functionscounting, weighing
InterfaceRS-232

Related product manuals