32 CFS/CCS-BA-e-1220
8.2 Linearisation CFS 50K-3 models
Operation
Display
Switch-on balance and during the selftest press
.
With the numerical keys enter the password „
9999“:
Confirm input with .
„tECH LoCAL“ is displayed, should it not be
displayed, with select „tECH LoCAL“.
Acknowledge with .
The appeared [] displays the current balance.
If necessary, at balance zero display using
select the weighing unit [kg or lb], which shall be
used for adjustment. The appeared [] displays the
current weighing unit.
Acknowledge with .
Ensure that there are no objects on the weighing
pan. After successful standstill control press the
button.
„LoAd 1“ will be displayed
15 kg adjustment weight placed carefully in the
centre of the weighing plate, wait for stability display
and confirm with .
„LoAd 2“ will be displayed,
30 kg adjustment weight placed carefully in the
centre of the weighing plate, wait for stability display
and confirm with .

Table of Contents

Other manuals for KERN CFS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the KERN CFS and is the answer not in the manual?

KERN CFS Specifications

General IconGeneral
Plate materialstainless steel
Readability0.01 g
Reproducibility0.01g
CalibrationExternal calibration
DisplayLCD
Weighing unitsg, lb, oz
Operating temperature10 °C - 40 °C
Functionscounting, weighing
InterfaceRS-232

Related product manuals