EasyManuals Logo

KERN CFS User Manual

KERN CFS
81 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #42 background imageLoading...
Page #42 background image
42 CFS/CCS-BA-e-1220
10.2 Automatic adding-up
With this function the individual weighing values are automatically added into the
summation memory when the balance is unloaded without pressing and edited,
when an optional printer is connected.
CFS 3K-5, CFS 6K0.1, CFS 15K0.2, CFS 30K0.5 models:
Menu settings:
F1 off ACC ON“ and „F2 Prt P mode Print Au ON“,
(see chap. 13.2)
When using as counting system there can be totalized on the reference as
well as on the bulk material balance.
Before the totalizing process select the active balance, see chap. 9.3.
CFS 50K-3 models:
Menu settings:
F1 off F2 Prt“ P mode Print Au on“,
see chap. 13.2
When using as counting system there can be totalized on the reference as
well as on the bulk material balance.
Before the totalizing process select the active balance, see chap. 9.3.
Add up:
Place weighing goods A.
After the standstill control sounds a signal tone. Unload the weighing good, the
weighing value is added into the summation memory and printed out.
Place goods to be weighed B.
After the standstill control sounds a signal tone. Unload the weighing good, the
weighing value is added into the summation memory and printed out.
Add more weighed goods as described before. Please note that the balance
must be unloaded between the individual weighing procedures.
This process can be repeated 99 times or until the weighing range of the balance
is exhausted.
Display and deletion of weighing data as well as print example (see chap.
10.1.1 and/or 10.1.2 (CFS 50K-3)

Table of Contents

Other manuals for KERN CFS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the KERN CFS and is the answer not in the manual?

KERN CFS Specifications

General IconGeneral
Plate materialstainless steel
Readability0.01 g
Reproducibility0.01g
CalibrationExternal calibration
DisplayLCD
Weighing unitsg, lb, oz
Operating temperature10 °C - 40 °C
Functionscounting, weighing
InterfaceRS-232

Related product manuals