EasyManuals Logo
Home>KERN>Scales>CFS

KERN CFS User Manual

KERN CFS
81 pages
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #81 background imageLoading...
Page #81 background image
CFS/CCS-BA-e-1220 81
18.1 Error messages
Error message
Description
Possible causes
Zeroing range exceeded
due to switching-on balance
or pressing (normally
4% max)
Object on the weighing
plate
Overload when zeroing
Inappropriate
adjustment
Damaged weighing cell
Damaged electronics
Keyboard error
Improper use of the
balance
Value outside the A/D
changer range
Weighing plate not
installed
Damaged weighing cell
Damaged electronics
Adjustment error
Inappropriate adjustment
Should other error messages occur, switch balance off and then on again. If the
error message remains inform manufacturer.

Table of Contents

Other manuals for KERN CFS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the KERN CFS and is the answer not in the manual?

KERN CFS Specifications

General IconGeneral
BrandKERN
ModelCFS
CategoryScales
LanguageEnglish

Related product manuals