EasyManuals Logo

KERN CFS User Manual

KERN CFS
81 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #75 background imageLoading...
Page #75 background image
CFS/CCS-BA-e-1220 75
Press , „div“ is displayed
Press , the currently set legibility is displayed,
Using select the desired readibility
Acknowledge by .
(example)
Press , „A2t“ is displayed
Automatic zero point correction (Autozero) by changing the
display, digits selectable (0.5d, 1d, 2d, 4d)
Press , the currently set value is displayed.
Press to select the desired setting (0.5d, 1d, 2d, 4d) and
acknowledge by .
Press , „0 mAnL“ is displayed
Zero setting range: Load range where the display is set to
zero by pressing .
Press , the currently set auto-zero range is displayed.
Press to select the desired setting (0 %, 2 %, 5 %, 10 %,
20 %) and acknowledge by pressing .
Press , „Pin“ is displayed
Password input for menu block „
Press , will be shown. Use the numeric
keyboard to enter the desired code and acknowledge with .
is displayed, request to repeat the password
input.
Enter code anew and press to acknowledge. If entry was
successful, is displayed, if false,
appears. In this case repeat the code input.

Table of Contents

Other manuals for KERN CFS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the KERN CFS and is the answer not in the manual?

KERN CFS Specifications

General IconGeneral
Plate materialstainless steel
Readability0.01 g
Reproducibility0.01g
CalibrationExternal calibration
DisplayLCD
Weighing unitsg, lb, oz
Operating temperature10 °C - 40 °C
Functionscounting, weighing
InterfaceRS-232