Remote Operations 6
Keysight 11713D/E Operating and Service Manual 79
HCOPy:SDUMp:DATA?
Provides screen shots of the front panel display. Returns an image of the display in .bmp
format.
HCOPy:SDUMp:DATA?
Examples
HCOP:SDUM:DATA?
Related Commands
DISPlay:ATTenuation
DISPlay:IO
DISPlay:SYSTem
DISPlay:CYCLe
DISPlay:INFormation
DISPlay:UTILity
SYSTem: ERRor?
Returns the next error number followed by its corresponding error message string from
the remote programming error queue. The queue is FIFO (first-in, first-out) buffer that
stores errors as they occur. As it is read, each error is removed from the queue. When all
errors have been read, the query returns “0,No Error”. If more errors are accumulated
than the queue can hold, the last error in the queue is “-350,Too Many Errors”.
SYSTem:ERRor?
Examples
To query the next error from the queue:
SYST:ERR?
Query? Yes, query only
Set? No
Return Parameters Binary Data
Minimum n/a
Maximum n/a
Units None
Query? Yes, query only
Set? No
Return Parameters Integer, String
Minimum n/a
Maximum n/a
Units None