Remote Operations 6
Keysight 11713D/E Operating and Service Manual 87
Related Commands
*CLS
*OPC?
*RST
*OPC?
Sets the standard event status register to monitor the completion of all commands. The
OPC command query returns the ASCII character “1” in the output queue to indicate
completion of all pending operations.
*OPC?
Examples
To query the Standard Event Status Register:
*OPC?
Related Commands
*CLS
*IDN?
*RST
*RST
Resets the switch/attenuator driver to the following factory-defined states:
- All channels in both banks will be lighted on.
*RST
Query? Yes, query only
Set? No
Return Parameters Integer, 0|1
Minimum n/a
Maximum n/a
Units None
Query? No
Set? Yes
Set Parameters n/a
Minimum n/a
Maximum n/a
Units None