6 Remote Operations
88 Keysight 11713D/E Operating and Service Manual
Examples
To reset the instrument:
*RST
Related Commands
*CLS
*IDN?
*OPC?
*SRE
Enables the bits in the status byte enable register. The selected enabled bits are
summarized in the “Master Status Summary” (MSS) bit (bit 6) of the Status Byte register.
If any of the selected bit condition changes from 0 to 1, a service request is generated.
Table below shows the contents of this register.
The query returns the current contents of the status byte enable register. The format of
the return is in the range of 0 to 255.
Table 6-2 Bit configuration of Status Byte Enable register
*SRE <Integer>
Examples
To set the bit in the status byte enable register to 128:
*SRE 128
Bit Position765432-0
Bit Name OPER
MSS
RQS
ESB MAV QUES n.u.
Bit Weight 128 64 32 16 8
n.u. - Not Used
OPER - Operation Status Summary
MSS - Master Status Summary
RQS - Request for Service
ESB - Event Status Byte Summary
MAV - Message Available
QUES - Questionable Status Summary
Query? Yes
Set? Yes
Set/Return Parameters Integer (register binary value)
Minimum 0
Maximum 255
Units None