Remote Operations 6
Keysight 11713D/E Operating and Service Manual 89
To query the status byte enable register:
*SRE?
Related Commands
*ESE
*ESR?
*STB?
*STB?
Reads the condition register of the Status Byte register and returns a decimal value
which corresponds to the binary-weighted sum of all bits set in the register.
*STB?
Examples
To query the condition register of the status byte register:
*STB?
Related Commands
*ESE
*ESR?
*SRE
*LRN?
Returns a binary string of all the saved EEPROM values required to set the instrument to
its current state.
The return string includes “SYSTem:SET” command header that can be used directly,
provided the receiver accepts binary data.
*LRN?
Query? Yes
Set? No
Return Parameters Integer (register value)
Minimum n/a
Maximum n/a
Units None
Query? Yes, query only
Set? No
Return Parameters (Optional) Binary Data
Minimum n/a
Maximum n/a
Units None