Operating Your Frequency Counter 2
Keysight 53150A/151A/152A Operating Guide 97
Power Correction Example: Editing Data Point Values
1 Select a power-correction profile using the up- and/or down-arrow keys (as
shown in steps 1 through 4 in the previous example). For this example, choose
profile number 3.
2 Press the right-arrow key.
The display changes to show the loss and frequency values stored in the data
point that contains the lowest frequency value in the current profile. The left
and right pointers that are before and after the word ā€œLOSSā€ are flashing.
The ten small, single-line annunciators that are below the analog power
display are used to indicate which of the ten data points is currently displayed.
In the example shown above, only the first annunciator is activated, which
indicates that the values displayed are for data point 1, the lowest data point.
dB