CALibrate:PASSword "<password>"
Sets the Admin password, which is used for calibration.
Parameter Typical Return
an alphanumeric string of 4 to 15 characters (none)
Set the password to 12345: CAL:PASS "12345"
The password must be 4 to 15 characters long and may include any of these characters:
Characters Description
A-Z Upper case letters
a-z Lower case letters
0-9 Numeric digits
+ - ( ) . , <space> Plus, minus, parentheses, period, comma, space
The CALibrate:STATe must be ON to execute this command.
This command increments the calibration count by 1.
The password is case sensitive.
To change the password: unsecure calibration memory with the old code, then set the new code.
If you enter an empty string as the new password and save it with CAL:SAVE once, no password will
be required to newly enter the calibration mode the next time.
This setting is non-volatile; it will not be changed by power cycling or *RST.
CALibrate:SAVE
Saves the calibration data and date, not the administration password, into nonvolatile memory. Do
this at the end of the calibration to avoid losing changes.
Parameter Typical Return
(none) (none)
Store calibration constants into nonvolatile memory: CAL:SAVE
The CALibrate:STATe must be ON to execute this command.
This command increments the calibration count by 1.