OUTPut Subsystem
The OUTPut subsystem configures the output state, power-on state, coupling mode, and protection.
OUTPut[:STATe] ON|1|OFF|0
OUTPut[:STATe]?
Enables or disables the instrument's output.
Parameter Typical Return
ON|1|OFF|0 0(OFF)or1(ON)
Turn on the output: OUTP ON
The *RST command sets this parameter to 0.
OUTPut:COUPling AC|DC|ACDC
OUTPut:COUPling?
Sets the output coupling mode.
Parameter Typical Return
AC|DC|ACDC
Default AC
AC, DC, or ACDC
Set the output mode to ACDC: OUTP:COUP ACDC
AC (default) produces ACoutput
DC produces DC output
ACDC combines ACand DCoutput
The *RST command sets this parameter to AC.