OUTPut:PON:STATe RST|RCL0|AUTO
OUTPut:PON:STATe?
Sets the power-on output state.
Parameter Typical Return
RST|RCL0|AUTO RST, RCL0, or AUTO
Set the power-on state to the *RST state: OUTP:PON:STAT RST
RST specifies that the instrument powers on in a state equivalent to *RST.
RCL0 loads the state stored in state storage memory location 0.
AUTO specifies that the instrument powers on in the power-down state. The output state is always off.
This parameter is not affected by *RST or *RCL.
This parameter is saved in nonvolatile memory.
OUTPut:PROTection:CLEar
Resets the latched protection. This clears the latched protection status that disables the output when
a protection condition occurs.
Parameter Typical Return
(none) (none)
Clears the latched protection status: OUTP:PROT:CLE
All conditions that generate the fault must be removed before the latched status can be cleared. The
output remains in the OFF state after the fault condition is cleared.
OUTPut:PROTection:WDOG[:STATe] ON|1|OFF|0
OUTPut:PROTection:WDOG[:STATe]?
Enables or disables the I/O watchdog timer.
Parameter Typical Return
0|OFF|1|ON 0(OFF)or1(ON)
Enables the watchdog timer protection: OUTP:PROT:WDOG ON
When the watchdog timer is enabled, the output will be disabled if there is no I/O activity on any
remote interface within the time period specified by OUTput:PROTection:WDOG:DELay. The output is
latched off but the programmed output state is not changed.
The watchdog timer function is NOT reset by front-panel activity; the output will still shut down after
the time period has elapsed.
The *RST command sets this parameter to 0.
3SCPI Programming Reference
142 Keysight AC6800B Series Operating and Service Guide