INITiate Subsystem
The INITiate commands initialize the trigger system. This moves the trigger system from the "idle"
state to the "wait-for-trigger" state, which enables the instrument to receive triggers. An event on the
selected trigger source causes the trigger to occur.
INITiate[:IMMediate]:ACQuire
Initiates the ACQuire (measurement) operation.
Parameter Typical Return
(none) (none)
Initiate a measurement: INIT:ACQ
If the ACQuire operation is already initiated, this command will produce error -213,"Init ignored".
INITiate[:IMMediate]:TRANsient
Initiates the TRANsient operation.
Parameter Typical Return
(none) (none)
Initiate a TRANsient operation:
VOLT:MODE STEP
FREQ:MODE STEP
INIT:TRAN
If the TRANsient operation is already initiated, this command will produce error -213,"Init ignored".
INITiate:CONTinuous:ACQuire ON|1|OFF|0
Initiates a new measurement and enables or disables "continuous" mode.
Parameter Typical Return
ON|1|OFF|0 0(OFF)or1(ON)
Initiate continuous acquisition mode: INIT:CONT:ACQ ON
If continuous mode is on, acquired data is automatically refreshed without the need for an INIT.
You should normally leave continuous mode off (default).
If the TRANsient operation is already initiated, this command will produce error -213,"Init ignored".
The *RST and *RCL commands set this parameter to 0.
Keysight AC6800B Series Operating and Service Guide 139
3SCPI Programming Reference