EasyManua.ls Logo

Keysight Technologies AC6801B - Calibration Procedure

Keysight Technologies AC6801B
197 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Calibration Procedure
Enter Calibration Mode
DC Voltage Low Range
DC Voltage High Range
AC Voltage Low Range
AC Voltage High Range
Current Low Range
Current High Range
End Calibration
Enter Calibration Mode
To begin the calibration process, log in using the Admin password. The factory default is no password.
Front Panel SCPI
Select System>Admin>Login. Then enter the password and press
[Select].
CAL:STAT ON
[,<password>]
DC Voltage Low Range
Connect the voltage input of the 3458A DMM to the instrument output as shown below and configure
the 3458Afor DCV measurements. You can use the following SCPI commands (or their front-panel
equivalents) to configure the DMM:
*RST
DCV AUTO, AZERO ON, FIXEDZ OFF
NPLC 100
Keysight AC6800B Series Operating and Service Guide 171
4Calibration, Verification, and Service

Table of Contents