End Calibration
This section describes the final steps in the calibration process.
Storing calibration constants overwrites the existing ones in nonvolatile memory. If you
do not save the calibration data, the new calibration constants will only be used until you
exit the calibration state or cycle power. For the front panel, leaving the calibration state
is defined as exiting the menu levels at or below [Menu]>System>Admin>Cal. For
SCPI, leaving the calibration state is accomplished by CAL:STAT OFF.
If you are not sure you want to permanently store the new constants, do not save the
data when you exit the calibration mode.
Step Description Front Panel SCPI
1 Enter the calibration
date.
[Menu]>System>Admin>Cal>Date
Enter the calibration date in the Data field. Press
[Select] or [Enter] to finish.
CAL:DATE"<
string>"
2 Save calibration data. [Menu]>System>Admin>Cal>Save CAL:SAVE
3 Log out. [Menu]>System>Admin>Logout CAL:STAT OFF