Step Action Normal Result
4 Program output to RANGE 310, VOLT 0, OUTP:COUP DC, OFFSET 128. Output voltage at 128 VDC, CV annunciator
on, output current near:
2 A for AC6801B
4 A for AC6802B
8 A for AC6803B
16 A for AC6804B
5 Record DVM reading and calculate current. Readings within specified dc current read-
back limits.
4îîCalibration, Verification, and Service
184 Keysight AC6800B Series Operating and Service Guide