Specifications
Hardware
• Spot, sweep, and more measurement capabilities
• Sampling (time domain) measurement from 100 s
• Pulse measurement from 500 s pulse width
• Accurate Quasi-Static Capacitance Voltage (QSCV) measurement with leakage
current compensation
B1514A Medium current source/monitor unit (MCSMU).
• Slots occupied: 1
• Range up to 30 V/1 A pulsed (0.1 A DC) with 4-quadrant operation
• Pulse measurement from 50 s pulse width with 2 s resolution
• Oscilloscope view for voltage/current waveform monitoring
• Minimum measurement resolution 10 pA/0.2 V
B1520A Multi-frequency capacitance measurement unit (MFCMU).
• Slots occupied: 1
• AC impedance measurement (C-V, C-f, C-t)
• 1 kHz to 5 MHz frequency range with minimum 1 mHz frequency resolution
• 25 V built-in DC bias and 100 V DC bias with SMU and SCUU (SMU CMU
Unify Unit)
• Easy and fast yet accurate IV and CV automated connection change by SCUU
B1525A High voltage semiconductor pulse generator unit (HV-SPGU).
• Slots occupied: 1
• High voltage output up to ± 40 V applicable for non-volatile memory testing
• Two-level and three-level pulse capability by single channel
• Flexible arbitrary waveform generation with 10 ns resolution (arbitrary linear
waveform generation function)
• Two channels per module
B1530A Waveform generator / fast measurement unit (WGFMU).
• Slots occupied: 1