2- 28 Keysight B1500A User’s Guide, Edition 14
Introduction
Measurement Modules
HRSMU - High Resolution SMU
This section describes typical specification of the high resolution source/monitor
unit (HRSMU). There is no restriction for the number of modules installed in one
mainframe.
Maximum voltage, current, output power: ± 100 V, ± 100 mA, 2 W
Minimum range: 0.5 V, 10 pA (1 pA when the E5288A ASU is used)
Output/measurement value and resolution: see Table 2-11 to Table 2-14.
Figure 2-6 HRSMU Output and Measurement Ranges