Keysight B1500A User’s Guide, Ed ition 14 4- 21
Specifications
Hardware
Overshoot (filter ON)
For HPSMU, MPSMU, HRSMU
Voltage source: 0.03 % of V range
Current source: 1 % of I range
For MCSMU
Voltage/Current source: 10 % of range
Range switching transient noise
Voltage ranging: 250 mV
Current ranging: 70 mV
Condition: filter ON for HPSMU, MPSMU, and HRSMU
Maximum guard offset voltage
± 1 mV (HPSMU)
± 3 mV (MPSMU, HRSMU)
± 4.2 mV (HRSMU/MPSMU with ASU, Iout 100 A)
Maximum slew rate
0.2 V/s (HPSMU, MPSMU, HRSMU)
1 V/s (MCSMU)
Maximum DC floating voltage
± 200 V DC between low force and common (MCSMU)