EasyManuals Logo

Keysight Technologies IntegraVision PA2201A Series User Manual

Keysight Technologies IntegraVision PA2201A Series
178 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #120 background imageLoading...
Page #120 background image
Equations for single-phase wiring groups
1P-2W-1M or 1P-2W-3M 1P-2W-3M only
Voltage
V RMS v= ( )
x x
( )
V V V= + / 2
avg 1 2
V RMS v v= ( )
diff 1 2
Current
I RMS i= ( )
x x
( )
I I I= + / 2
avg 1 2
I RMS i i= ( + )
n 1 2
Power
P AVG v i= ( * )
x x x
P P P= +
sum 1 2
Apparent
Power
S V I= *
x x x
Reactive
Power*
Q S P= ±
x x x
2 2
Q Q Q= +
sum 1 2
Power Factor
λ P S= /
x x x
λ P S= /
sum sum sum
Power Angle
a λcos( )
x
Rectified Mean
Voltage
V AVG v= ( )
rect x
Rectified Mean
Current
I AVG i= ( )
rect x
Crest Voltage
V MAX V V RMS v= ( , ) / ( )
crest xmax min
Crest Current
I MAX I I RMS i= ( , ) / ( )
crest xmax min
* For Reactive Power, the sign in front of the square root is + for lagging current and -1 for
leading current. The Voltage and Current CWA frequency must be within 5% to detect leading
or lagging.
118 Keysight IntegraVision PA2200 Series Power Analyzers Operating and Service Guide

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies IntegraVision PA2201A Series and is the answer not in the manual?

Keysight Technologies IntegraVision PA2201A Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelIntegraVision PA2201A Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals