437 Keysight CXG, EXG, and MXG X-Series Signal Generators Service Guide
Functional Tests
Leveled Output Power
15.Turn the arbitrary waveform player on: Press the ARB softkey to highlight
On.
16.Set the frequency to the first value listed in Table 17-4:
Press Frequency > 50 > MHz.
17.Set the amplitude to 7 dBm:
Press Amplitude > 7 > dBm.
18.If using a power meter, configure the power meter as follows:
a. On the power meter, press the Frequency Cal Fac button.
b. Select a power meter channel (if applicable).
c. Use the arrow keys to enter the frequency at which to measure the
power.
19.Measure the output power.
20.Repeat steps 16 through 19 for the remaining frequencies listed in Table
17-4, and confirm that the power level at each point is within limits.
Table 17-4 Output Power With Modulation
Output Power With Modulation
Frequency (MHz) Amplitude (dBm) Limits (dB)
50 7 ±1.05
100 7 ±1.05
250 7 ±1.05
323 7 ±1.05
512 7 ±1.05
814 7 ±1.05
1275 7 ±1.05
2025 7 ±1.05
2750 7 ±1.05
3750 7 ±1.05
5250 7 ±1.05

Table of Contents

Other manuals for Keysight Technologies N5173B EXG

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies N5173B EXG and is the answer not in the manual?

Keysight Technologies N5173B EXG Specifications

General IconGeneral
BrandKeysight Technologies
ModelN5173B EXG
CategoryPortable Generator
LanguageEnglish

Related product manuals