EasyManua.ls Logo

Keysight Technologies N5183B MXG - Storing, Loading, and Playing a Waveform Segment

Keysight Technologies N5183B MXG
451 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Keysight CXG, EXG, and MXG X-Series Signal Generators Service Guide 155
Troubleshooting
Pulse Modulation Issues (Option UNW)
4. Set the signal level to 0 dBm by pressing AMPTD, 0 dBm.
5. Set the Pulse Width to 2 µsec by pressing Pulse, Pulse Width, 2 µsec.
6. Set the Pulse Period to 4 µsec by pressing Pulse Period, 4 µsec.
7. Turn pulse modulation on by setting Pulse to On.
8. Connect the RF Output to a spectrum analyzer with the following settings:
Center Frequency = Signal Generator Frequency
Span = 0 Hz
Reference Level = 0.0 dBm
Sweep Time = 20.0 µs
If the pulse modulation was working properly there would be a pulsed
signal on the spectrum analyzer with a peak amplitude of approximately
0 dB.
9. Referring to Figure 6-10, connect an oscilloscope to U77 pin 2 on the
A7A1 Microwave ALC Control board assembly.
Figure 6-10 Pulse Modulation Drive Location – A7A1 U77 Pin 2
10.Setup the oscilloscope as follows:
Amplitude scale = 500 mv/div
Horizontal scale = 2 µs/div
11.A signal like that shown in Figure 6-11 should be seen on the oscilloscope
screen.

Table of Contents

Other manuals for Keysight Technologies N5183B MXG

Related product manuals