Keysight N9927-90001 User’s Guide 181
SA (Spectrum Analyzer) Mode
SA Mode Settings
RF Burst Amplitude Alignment
Channel Equalization
These are referred to as Individual Align
.
All three or a subset of the individual alignments are required for each specific
mode, depending on the type of measurement supported by it:
SA:
InstAlign Amplitude Alignment
RF Burst Amplitude Alignment
RTSA:
InstAlign Amplitude Alignment
RF Burst Amplitude Alignment
Channel Equalization
Channel Scanner:
InstAlign Amplitude Alignment
ERTA:
InstAlign Amplitude Alignment
IQA:
InstAlign Amplitude Alignment
RF Burst Amplitude Alignment
Channel Equalization
Built-in Power meter
InstAlign Amplitude Alignment
Default settings and recommended use model for many applications:
Coupled individual alignments
—Coupled state is Auto
Individual alignments are not accessible via front panel soft-keys
Individual alignments cannot be UnCoupled
For special cases, the individual alignments can be accessible via the front
panel soft-keys using the Preference setting SA - Individual Alignment
Control
> On/Off. This enables a soft-key for access to Individual Alignments
when this preference is set to ON: Cal 5
> Individual Align. For more refer to
Preferences on page 515, in Chapter 26 System Settings.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight FieldFox A Series and is the answer not in the manual?

Keysight FieldFox A Series Specifications

General IconGeneral
BrandKeysight
ModelFieldFox A Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals